To generate binary T-x-y data using Ebulliometer (vle)
Purge the System with Nitrogen gas to remove contaminants.
- Open the outlet of Raal’s apparatus.
- Start Nitrogen Gas.
- Allow gas to flow through the system for few minutes.
- Shut off nitrogen supply after pop up massage for purging completion is displayed.
Generating boiling point data at Atmospheric pressure
- Select the system and composition from the dropdown list.
- Close the outlet of Raal’s apparatus and open the inlet.
- Close the inlet of Raal’s apparatus after it is filled.
- Start cold water pump.
- Start Heating coil.
- Increase heating rate slowly by moving the sliding bar.
- Allow system to accommodate change in the heat.
- Observe the change in the temperature.
- Do the same until small increment in heating rate won’t rising temperature.
- Take the liquid and vapor samples.
- Repeat the procedure for different compositions.
Sample Analysis.
- Make 4 samples containing known mass of individual components of system + external standard (ES).
- Calibrate GC with this sample and note down the area for each.
- Plot the graph of mass ratio of component to ES Vs area ratio of component to ES.
- Calculate slope.
- Analyze samples using GC.
- Use slope and basic information of individual compounds (i.e. molecular wt., Density,…) to calculate mole fraction.