To determine the infinite dilution activity coefficients (IDACs) of a binary mixture using Ebulliometer
Purge the System with Nitrogen gas to remove contaminants.
- step 1. Open the outlet of Raal’s apparatus.
- step 2. Start Nitrogen Gas.
- step 3. Allow gas to flow through the system for few minutes.
- step 4. Shut off nitrogen supply after pop up massage for purging completion is
displayed.
Generating boiling point data at Atmospheric pressure.
- step 1. Select the system and composition from the dropdown list.
- step 2. Close the outlet of Raal’s apparatus and open the inlet.
- step 3. Close the inlet of Raal’s apparatus after it is filled.
- step 4. Start cold water pump.
- step 5. Start Heating coil.
- step 6. Increase heating rate slowly by moving the sliding bar.
- step 7. Allow system to accommodate change in the heat.
- step 8. Observe the change in the temperature.
- step 9. Do the same until small increment in heating rate won’t rising temperature.
- step 10. Take the liquid and vapor samples.
- step 11. Repeat the procedure for different compositions.
Sample Analysis.
- step 1. Make 4 samples containing known mass of individual components of system
+ external standard (ES).
- step 2. Calibrate GC with this sample and note down the area for each.
- step 3. Plot the graph of mass ratio of component to ES Vs area ratio of component
to ES.
- step 4. Calculate slope.
- step 5. Analyze samples using GC.
- step 6. Use slope and basic information of individual compounds (i.e. molecular wt.,
Density,…) to calculate mole fraction.